IMMERSIBLE HEAD GCZ-1t


The GCZ-1t immersible head is designed to cooperate with controllers used for making continuous measurements of conductivity in distilled water, tap water, and waste water in sewage tanks or drains.
In the bottom part of the head the conductivity and temperature sensors are placed.
In the closed part the preamplifier is placed.
In the upper part there is a connector for the cable connecting with the measuring device.
In our offer we have two kinds of heads which differ in measuring ranges:
1. 0,02 mS/cm ÷ 200 mS/cm
, 2. 0 μS/cm ÷ 9999 μS/cm
. The conductivity sensor is equipped with stainless steel electrodes.
Periodical calibration is simple and proceeds after pulling the head out of the measured liquid and with use of the plexiglass cover into which the standard solution is pourred. This cover is also used for periodical cleaning of the cell's electrodes.
The head is equipped with a bow to be hanged on an outrigger.
The standard length of the whole head is 1m. It is possible to prepare a head with length from 0.5m to 4m. The head cooperates exclusively with the meters of 800 series. The signal is amplified by a preamplifier integrated with the cell.



Technical Data
Length with the bow External diameter Minimal immersion Material
standard 1m. maximal 4m. 50 mm 50 mm PVC