OPTICAL OXYGEN SENSOR DO-1


The optical oxygen sensor is designed for measurements of concentration of oxygen dissolved in liquid. It has several advantages in comparison with membrane, galvanic sensors.

Characteristic features:



The sensor cooperates exclusively with CO-404 meter adjusted to optical measurements.

Operation

The sensor works basing on changes of activity of the fluorescent dye, which covers the replaceable luminescent cap. The fluorescence reduction affected by light is dependent on oxygen concentration in the tested liquid. The higher concentration, the faster fluorescence reduction. This dependence is calculated with use of an appropriate formula. DO-1 is equipped with built-in temperature sensor used for automatic temperature compensation.

Accessories

The sensor is equipped with a calibration vessel used for calibration and storage in optimally humid environment.


TECHNICAL DATA
Oxygen measurement Range 0 ÷ 20 mg/l ; 0 ÷ 200 %
Accuracy ± 0.30 mg/l
Stabilisation time (T90) 20 s.
Offset < 0,10 mg/l
Repeatability < 0,15 mg/l
Drift 5 % / year
Working time of the luminescent cap ≥8000 hours
Temperature measurement Range 0 ÷ 50 °C
Accuracy ± 0,5 °C
Communication protocol RS-485
Cable length 3 m.
Sensor's dimensions φ = 16 mm, L = 105 mm