TEMPERATURE SENSOR CT1U-146 FOR PT-401
The temperature sensor type CT1U-146 is designed for temperature measurements in liquid and semi-solid substances and also for air and gases measurements. It may be used during the laboratory, field and industrial measurements. The measuring element is a platinum resistor Pt-100 Class 1/10B made by Heraeus. The sensor has an acidresistant stainless steel body which in most cases protects the measuring element against mechanical damage and against damage caused by aggressive solutions. As standard the sensor is equipped with a 1m silicone cable and 8-pin connector. The silicone cable is resistant to temperatures up to 250 °C, it may also be placed in a steel braid to protect it against mechanical damage. The end of the sensor may be flat or sharp, depending on the customer's needs. The error of the Pt-100 sensor is changing together with the temperature change. IN 0°C are the lowest . The maximal error of the 1/10 B clas sensor is calculated from the formula: dtmax = ±(0.03 °/C + 0.0005t) , where: t – displayed temperature in °C. The total error of the measurement is a sum of the sensors and the meters errors.
Technical data
Measuring range | -50 ÷ 300 °C |
---|---|
Measuring element | Pt-100 Class 1/10B |
Error in the 0 °C | 0.03 °C |
Error in the range 0 ÷ 100 °C | 0.08 °C |
Body diameter | 4.0 mm ± 0.5 mm |
Body length | 140 mm |
Minimal depth of immersing | 10 mm |
Reaction time t 99 from 20 to 100 °C | 11 s |
Body material | stainless steel type 0H18N9 |
Cable length | about 1 m |
Cable material | silicone, resistant to 250 °C |
Handle material | polyamide, resistant to 200 °C |
Connector | 8 pin |