pH / ION METER CPI-601
The most important features
- Very high accuracy meter for laboratory use.
- Measures: pH, redox potential, ion concentration in pX, g/l, M/l, or ppm and temperature.
- All the results are simultaneously displayed on a 7” graphical colour touch screen.
- Calibration in 1 to 5 points.
- The meter has separate BNC connectors for the pH electrode and the ion selective (or redox) electrode, and a banana connector for the reference electrode.
- In the pH function, automatic correction of the pH standard value with temperature (NIST standards).
- In the ION measurement function, molar weight and valence of the measured ion are automatically entered after choosing the ion.
- Automatic conversion of units – e.g. M/l to mg/l.
- Memory of 2000 data sets.
- USB output.
- Meets the GLP requirements.
Detailed description
CPI-601 measures: pH, redox (mV), ion concentration and temperature. The meter enables simultaneous measurement and display of the ion selective reading for the chosen ion, pH (or mV) and temperature. The results are displayed on a 7” graphical colour touch screen.Characteristic features:
- Programming of the parameters is very easy.
- Unification of operating procedures for all functions makes working easier.
- “HOLD” function to freeze the result on the display.
- Signalisation of the result stabilisation with the “READY” symbol and a sound.
- Possibility of sending a calibration report to a PC - up to 10 last calibrations.
In the pH measuring function:
- pH electrode calibration in 1 to 5 points.
- Automatic detection of pH buffer and standard values entered by the user.
- When NIST pH standards are used, their stored values are automatically corrected for temperature, so there is no need to heat or cool the solutions.
- Storing the characteristics of 3 pH electrodes enables their quick replacement.
- Automatic control of the electrode’s condition.
- Readout of the pH electrode parameters – zero offset and slope.
- Depending on the electrode used, measurements in pure water, sewage, pastes, etc. are possible.
- The measuring circuits of the pH and ion connectors are isolated, which enables accurate, error-free simultaneous measurements in the same vessel.
In the ion selective measuring function:
- Enables ion concentration measurements of monovalent, bivalent, negative and positive ions.
- The measuring range enables work with all available ion selective electrodes (ISE), chosen adequately to the kind of measured ion, equipped with BNC-50 connector.
- The meter automatically introduces the molar weight and valence of a particular ion.
- The user may choose the unit: pX, g/l, M/l, ppm.
- Automatic conversion of units – e.g. M/l to mg/l.
- Possibility of entering freely chosen values of ion standard solutions.
- The meter has separate BNC connectors for the pH electrode and the ion selective (or redox) electrode, and a banana connector for the reference electrode.
In the redox (mV) measuring function:
- Precise redox potential measurement (accuracy 0.1 mV).
- The mV measurement is possible using the pH or the ISE connector.
- Possibility to measure redox potential relatively to the entered or measured reference potential – Vref.
In the temperature measuring function:
- Choice of the unit: °C, °F, K.
- Entering the group number of the selected temperature sensor, which increases the accuracy.
Other features:
- Automatic or manual temperature compensation.
- Internal clock with date.
- Datalogger for 2000 data sets of all currently measured functions.
- Storing of measurement results with time and date, taken singly or in series at a set time interval.
- The results and calibration data are stored in non-volatile memory.
- Possibility of screen brightness control depending on the external conditions.
- Storing of the next calibration date.
- USB output to connect with a PC.
- Possibility to choose English or German language.
- IP64 ingress protection.
- The meter meets the GLP requirements.
- 24 months of warranty for the meter.
TECHNICAL DATA
| Function | pH | Redox / mV | Temperature |
|---|---|---|---|
| Range | -6.000 ÷ 20.000 pH | ±2000.0 mV | -50.0 ÷ 200.0 °C |
| Resolution | 0.001 pH or 0.01 pH | 0.1 mV | 0.1 °C |
| Accuracy (±1 digit) | ±0.002 pH* | ±0.1 mV* | ±0.1 °C** |
| Temperature compensation | -5 ÷ 110 °C | - | - |
| Input impedance | >1012Ω | >1012Ω | - |
| Power supply | 15 V / 1 A power adapter | ||
| Weight | 520 g | ||
| Dimensions (mm) | L = 188.5, W = 134, H = 58 (at the highest point) | ||
| * | The accuracy of the meter only. |
| ** | The accuracy of the meter only. The total error includes the meter’s and probe’s accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000S resistor: ±0.1 °C. |
| Function | Ion (M/l) | Ion (g/l) | Ion (ppm) | Ion (pX) |
|---|---|---|---|---|
| Range | 0 ÷ 100 | 0 ÷ 1 000 | 0 ÷ 1 000 000 | -2.000 ÷ 16.000 pX |
| Resolution | 0.01 / 0.1 % | 0.01 / 0.1 % | 0.01 / 0.1 % | 0.001 / 0.01 |
| Accuracy (±1 digit) | ±0.25 %* | ±0.25 %* | ±0.25 %* | ±0.002 pX* |
| Temperature compensation | -5 ÷ 110 °C | -5 ÷ 110 °C | -5 ÷ 110 °C | -5 ÷ 110 °C |
| * | The accuracy of the meter only. |
Subranges / resolution
| 1 | 0.00 µg/l ÷ 99.99 µg/l (µM/l, ppb) |
|---|---|
| 2 | 100.0 µg/l ÷ 999.9 µg/l (µM/l, ppb) |
| 3 | 1.000 mg/l ÷ 9.999 mg/l (mM/l, ppm) |
| 4 | 10.00 mg/l ÷ 99.99 mg/l (mM/l, ppm) |
| 5 | 100.0 mg/l ÷ 999.9 mg/l (mM/l, ppm) |
| 6 | 1.000 g/l ÷ 9.999 g/l (M/l, ppt) |
| 7 | 10.00 g/l ÷ 99.99 g/l (M/l, ppt) |
| 8 | 100.0 g/l ÷ 999.9 g/l (M/l, ppt) |