pH / CONDUCTIVITY METER CPC-811
Detailed description:
CPC-811 is used for continuous measurement of pH, conductivity and temperature. May be used in agriculture for controlling parameters of the liquids during drip irrigation, it may also be used for measurements in aquariums.Characteristic features:
- Waterproof housing (IP-65) enables work in difficult conditions.
- 2 displays enable simultaneous observation o the pH, conductivity and temperature measurement results.
- Isolated pH and conductivity circuits enable measurements in this same vessel or tank without any problems.
- After choosing a right type of measuring head, with electrodes, may be used for measurements in pipelines or tanks.
- Very easy in use.
- Powered with power adapter.
In the pH measurement mode:
- Calibration in 1 ÷ 3 points in buffer solutions.
- Automatic detection of the pH buffer solution’s value 4.00, 7.00, 9.00 pH (optionally: 4.00; 7.00; 10.00 pH).
- Automatic temperature compensation.
- Delivered with good quality plastic body pH electrode.
In the conductivity measurement mode:
- Enables measurements of samples with conductivity up to 100 mS/cm.
- 5 sub-ranges switched automatically (autorange).
- Automatic calibration in 1.41 mS/cm or 12.90 mS/cm solutions.
- The meter automatically detects the value of the used standard solution.
- Automatic temperature compensation.
- Constant temperature compensation coefficient α (2 %/°C).
- Constant reference temperature value 25 °C .
- Delivered with good quality conductivity cell.
Other features:
The standard set includes CT2B-121 temperature sensor with Pt-1000B resistor, pH electrode and conductivity cell chosen separately.
TECHNCAL DATA
Function | pH | Conductivity | Temperature | |
---|---|---|---|---|
Range | 0.00 ÷ 14.00 pH |
0 ÷ 100 mS/cm (autorange) |
-50.0 ÷ 199.9 °C | |
Accuracy (±1 digit) | ±0.01 pH* |
±0.25%* |
±0.1 °C** | |
Temperature compensation | -5.0 ÷ 110.0 °C | -5.0 ÷ 70.0 °C | - | |
Input impedance | >1012Ω | - | - | |
Temperature sensor | Pt-1000 standard or accurate | |||
Power supply | 12 V / 100 mA power adapter | |||
Weight | 470 g | |||
Dimensions (mm) | L = 160, W = 160, H = 60 |
* | The accuracy of the meter only. |
** | The accuracy of the meter only. The total error includes the meters and probe’s accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8 °C, with Pt-1000A resistor: ±0.35 °C. |