LABORATORY pH / CONDUCTIVITY / SALINITY METER CPC-511
The most important features
- Simple laboratory meter.
- Measures pH, redox potential (ORP), conductivity, salinity, TDS and temperature.
- The display enables simultaneous readout of the measured function and temperature.
- pH electrode calibration in 1 to 3 constant points (4.00, 7.00, 9.00 pH), conductivity in 1 point.
- In the conductivity measurement function constant value of the α coefficient (2%/°C).
- Memory of 50 results.
- No PC connection.
Detailed description
Easy in use laboratory measuring device, which measures pH, mV (redox potential), conductivity, salinity, TDS and temperature.Characteristic features:
- Large LCD enables simultaneous readout of the measured function and temperature.
- Very easy in use.
In the pH measuring function
- Calibration in 1 ÷ 3 points in buffer solutions.
- Automatic detection of the pH buffer solution’s value 4.00, 7.00, 9.00 pH (optionally: 4.00; 7.00; 10.00 pH).
- Meter shows information about the pH electrode condition (readout of the buffer and slope values).
- Automatic temperature compensation.
- Depending on the chosen electrode, makes measurement in clean water, sewage, pastes, etc.
In the conductivity measuring function
- Measures conductivity in distilled water and other liquids up to 1000 mS/cm.
- 5 sub ranges switched automatically (autorange).
- Salinity measurement converted to NaCl up to 250 g/l or KCl up to 200 g/l.
- Converts conductivity into salinity according to real characteristic and not a constant coefficient.
- Determines estimated value of the total dissolved solids (TDS).
- Calibration, by entering the K constant of the cell or by determining it with use of a standard solution.
- Possibility of entering the reference temperature value.
- Automatic temperature compensation.
- Constant α temperature coefficient 2 %/°C. For precise measurements with temperature compensation it is recommended to use meters with variable coefficient (CPC-502, CPC-505).
- Memory of 50 results.
- Wide temperature measuring range.
- Warranty for the meter: 24 months.
The set includes CT2B-121 temperature probe with Pt-1000B resistor, accurate ECF-1 conductivity cell and EPS-1 pH electrode for measurements in clean water. It should not be used in other types of liquids. Measurements in liquids with sediments should be made with use of IJ-44A pH electrode. Its unusual construction ”intermediate junction” protects the real junction (diaphragma) of the electrode against clogging, ensures stable measurements in these types of liquids or semi-liquid masses, in which other electrodes stop working quickly. When properly handled and maintained the electrode's lifetime is longer than the standard electrodes.
TECHNICAL DATA
Function | pH | mV |
Conductivity/ Salinity |
Temperature |
---|---|---|---|---|
Range | -2.00 ÷ 16.00 pH | ±1999 mV |
0 ÷ 1000 mS/cm (autorange) / 0 ÷ 200 g/l KCl 0 ÷ 250 g/l NaCl |
-50.0 ÷ 199.9°C |
Accuracy (±1 digit) | ±0.01 pH* | ±1 mV* |
±0.25%* / salinity 2 %* |
±0.1 °C** |
Temperature compensation | -5.0 ÷ 110.0 °C | - | -5.0 ÷ 70.0 °C | - |
Input impedance | >1012Ω | >1012Ω | - | - |
α coefficient | - | - | 2%/°C constant | - |
K constant | - | - | 0.010 ÷ 19.999 cm-1 | - |
Temperature sensor | Pt-1000 standard or accurate | |||
Power supply | 12 V / 100 mA power adapter | |||
Weight | 520 g | |||
Dimensions (mm) | L = 200, W = 180, H = 20/50 |
* | The accuracy of the meter only. |
** | The accuracy of the meter only. The total error includes the meters and probe’s accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8& deg;C, with Pt-1000A resistor: ±0.35 °C. |