WATERPROOF pH / CONDUCTIVITY / SALINITY METER
CPC-411


The most important features



Detailed description


Easy in use simple meter of good accuracy.
Measures: pH, mV (redox potential), conductivity, salinity and temperature.

Characteristic features:



In the pH measurement mode:


In the redox (mV) measuring function:


In the conductivity measurement mode:


Other features:


The standard set includes: ECF-1 conductivity sensor, CT2B-121 temperature sensor with Pt-1000B resistor, EPS-1 pH electrode for measurements in clean water. It should not be used in other types of liquid. Measurements in liquids with sediments should be made with use of IJ-44A pH electrode, which enables measurements in various samples of both pure and contaminated liquids and semi-solids. Its unusual construction (”intermediate junction”) protects the real junction (diaphragma) of the electrode against clogging, ensures stable measurements in these types of liquids or semi-liquid masses, in which other electrodes stop working quickly. When properly handled and maintained the electrode's lifetime is longer than the standard electrodes.



TECHNICAL DATA
Function pH mV Conductivity /
Salinity
Temperature
Range 0.00 ÷ 14.00 pH ±1999 mV 0 ÷ 199.9 mS/cm
(autorange) /
NaCl 0 ÷ 177.9 g/l
-50.0 ÷ 199.9 °C
Accuracy (±1 digit) ±0.01 pH* ±1 mV* ±0.25 %* /
salinity 2 %*
±0.1 °C**
Temperature compensation -5.0 ÷ 110,0 °C - -5.0 ÷ 70.0 °C -
Input impedance >1012Ω >1012Ω - -
α coefficient - - 2%/°C constant -
K constant - - 0.010 ÷ 1.999 cm-1 -
Temperature sensor Pt-1000 standard or accurate
Power supply 9 V battery, 12 V / 100 mA power adapter (option)
Weight 220 g
Dimensions (mm) L = 149, W = 82, H = 22
* The accuracy of the meter only.
** The accuracy of the meter only. The total error includes the meters and probe’s accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8 °C, with Pt-1000A resistor: ±0.35 °C.