pH-METER CP–661
The most important features
- pH meter for use in laboratory.
- Measures pH, redox potential (ORP) and temperature.
- Simultaneous observation of all the chosen functions results on the colour touch sreen.
- Calibration in 1 to 3 points.
- In the pH function automatic correction of the pH standard solution value changes, along with the temperature changes for NIST standards.
- Memory of 500 data sets.
- USB output.
- Small size and weight.
- The meter meets the GLP requirements.
Detailed description
CP-661 Is distinguished by a large a 5” colour graphic touch screen.The meter can be used for accurate measurements of pH, mV and temperature.
This pH meter enables simultaneous measurements displaying and storing of 2 functions (pH and temperature).
Characteristic features:
- Depending on the kind of applied electrode it may be used for clean water, sewage, soil, pastes, meat, cheese and other measurements.
- Calibration of the pH electrode in 1 ÷ 3 points with possibility of choosing the buffers value in a given range.
- Range of the entered calibration points:
1 point 0.000 ÷ 6.000 pH,
2 point 6.800 ÷ 7.100 pH,
3 point 8.000 ÷ 14.000 pH. - Automatic detection of buffer solutions and standard solutions entered in the meters memory.
- Automatic correction of the pH standard solution value along with the temperature changes for NIST standards, what eliminates the necessity of the standard solution temperature adjustment.
- The NIST standard solutions entered in the meters memory are phtalate, phosphate and borax or bicarbonate (chosen when the meter is ordered). The solutions in 20°C have following values: 4.001 pH, 6.881 pH, 9.225 pH or 10.064 pH.
- Possibility of storing characteristics of 3 electrodes enables their quick replacement, useful when using different types of electrodes specific for different types of samples.
- Automatic control of the electrode’s condition.
- Possibility of viewing the electrode's parameters (buffer and slope).
- Precise redox potential measurement (accuracy 0.1mV).
- The pH and conductivity measurement circuits are isolated, so there is no interference during simultaneous measurements.
- Automatic or manual temperature compensation.
Other features:
- Internal clock with date.
- Collecting up to 500 data sets in the internal data-logger with temperature, time and date, single collecting and taking series of measurements of all measured functions.
- Non-volatile memory of the stored results and calibration data.
- HOLD function to freeze the actual measurements results on the screen.
- Stabilised reading signalisation.
- Storing the calibration validity date and signalising it to the user.
- Possibility of choosing the language of the displayed information : English, German, French, Italian, Spanish, Portuguese.
- Possibility of connecting with a PC by a USB connector.
- Software for data transmission and collection delivered in set.
- Up to 10 last Calibration Reports transfer to a PC for printout.
- Powered with 5V/1000mA USB power adapter.
- The meter meets the GLP requirements.
- 24 months of warranty for the meter
The standard set includes CT2S-121 temperature sensor with Pt-1000S resistor, EPS-1 pH electrode for measurements in clear water. It should not be used in other types of liquid. Measurements in liquid with sediment should be made with use of IJ-44A pH electrode, which enables measurements in various samples of both pure and contaminated liquids and semi-solids. Its unusual construction (”intermediate junction”) protects the real junction (diaphragma) of the electrode against clogging, ensures stable measurements in these types of liquids or semi-liquid masses, in which other electrodes stop working quickly. When properly handled and maintained, the electrode's lifetime is longer than the standard electrodes.
TECHNICAL DATA
Function | pH | Redox / mV | Temperature |
---|---|---|---|
Range | -6.000 ÷ 20.000 pH | ± 2000 mV | -50.0 ÷ 200.0 °C |
Resolution | 0.001 pH or 0.01 pH | 0.1 mV | 0.1 °C |
Accuracy (±1 digit) | ±0.005 pH* | ±0.2 mV* | ±0.1 °C** |
Temp. compensation | -5 ÷ 110 °C | - | - |
Input impedance | >1012Ω | >1012Ω | - |
Power supply | Adapter 5 V / 1000 mA | ||
Weight | 420 g | ||
Dimensions (mm) | L = 175, W = 140, H = 52 |
* | The accuracy of the meter only. |
** | The accuracy of the meter only. The total error includes the meters and probe’s accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000S resistor: ±0.27 °C. |