LABORATORY pH METER CP-511
The most important features
- Simple laboratory meter.
- Measures pH, redox potential (ORP) and temperature.
- The display enables simultaneous readout of the measured function and temperature.
- pH electrode calibration in 1 to 3 constant points (4.00, 7.00, 9.00 pH).
- Automatic detection of the pH buffer solution value.
- Memory of 50 results.
- No PC connection.
Detailed description
Simple, easy-to-use laboratory meter for pH, mV (redox potential) and temperature measurements.Characteristic features:
- Large, easy-to-read LCD display.
- Simultaneous readout of the measured function and temperature value.
- Calibration of the pH electrode: 1 ÷ 3 points 4.00, 7.00, 9.00 or optionally 4.00, 7.00, 10.00 pH.
- Automatic buffer detection.
- Automatic temperature compensation.
- Depending on the pH electrode used, measurements of clean water, waste water, meat, cheese, pastes, soil etc. are possible.
- Possibility to check the electrode condition (buffer and slope).
- Memory of 50 results.
- Wide temperature measurement range.
- IP64 ingress protection.
- 24 months of warranty for the meter.
The standard set includes CT2B-121 temperature probe with Pt-1000B resistor and EPS-1 pH electrode for measurements in clean water. It should not be used in other types of liquids. Measurements in liquids with sediments should be made with use of IJ44A pH electrode. Its unusual construction (“intermediate junction”) protects the real junction (diaphragm) of the electrode against clogging, ensures stable measurements in these types of liquids or semi-liquid masses, in which other electrodes stop working quickly. When properly handled and maintained, the electrode's lifetime is longer than the standard electrodes.
TECHNICAL DATA
| Function | pH | Redox / mV | Temperature |
|---|---|---|---|
| Range | -2.00 ÷ 16.00 pH | ± 1999 mV | -50.0 ÷ 199.9 °C |
| Resolution | 0.01 pH | 1 mV | 0.1 °C |
| Accuracy (±1 digit) | ±0.01 pH* | ±1 mV* | ±0.1 °C** |
| Temp. compensation | -5 ÷ 110 °C | - | - |
| Input impedance | >1012Ω | >1012Ω | - |
| Power supply | 12 V / 100 mA power adapter | ||
| Weight | 600 g | ||
| Dimensions (mm) | L = 200, W = 180, H = 20/50 | ||
| * | The accuracy of the meter only. |
| ** | The accuracy of the meter only. The total error includes the meter’s and probe’s accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8 °C, with Pt-1000A resistor: ±0.35 °C. |