WATERPROOF pH-METER CP-451
The most important features
- Simple field and laboratory meter.
- Measures pH, redox potential (ORP) and temperature.
- Colour, graphic touchscreen - enables simultaneous readout of the pH or mV and temperature value.
- Waterproof housing (IP-66) enables work in difficult conditions.
- Calibration of the pH electrode in 1 to 3 points (pH: 4.00, 7.00, 9.00).
- Readout of the electrode's slope and offset.
- Memory of 200 results with time and date
- Favourable price
Detailed description
Used for pH, mV (redox potential) and temperature measurements.Characteristic features:
- Colour, graphic touchscreen - enables simultaneous readout of the pH or mV and temperature value.
- Easy-in-use.
- Waterproof housing (IP-66) enables work in difficult conditions.
- Small size and low weight make working in the field easier.
- Long lasting work in laboratory possible with use of the power adapter.
- Calibration of the pH electrode in 1 to 3 points (pH: 4.00, 7.00, 9.00).
- Automatic buffer detection.
- Automatic temperature compensation.
- Calibration time reminding.
- Readout of the electrode's slope and offset.
- Memory of 200 results with time and date
- Available 7 different languages: english, german, french, spanish, portuguese, italian, polish
- Powered by rechargeable batteries or power adapter.
- 24 months of warranty for the meter.
- The standard set includes the meter, the pH electrode, power adapter and a protective case.
IJ44A pH electrode
TECHNICAL DATA
Function | pH | Redox / mV | Temperature |
---|---|---|---|
Range | 0.00 ÷ 14.00 pH | ± 2000 mV | -50.0 ÷ 200.0 °C |
Resolution | 0.01 pH | 1 mV | 0.1 °C |
Accuracy (±1digit) | ±0.01 pH* | ±1 mV* | ±0.1 °C** |
Temp. compensation range | -5 ÷ 110 °C | - | - |
Input impedance | >1012Ω | >1012Ω | - |
Power | 2 x AA 1.2 V rechargeable battery, 5 V / 1000 mA power adapter | ||
Weight | 240 g | ||
Dimensions (mm) | L=149, W=82, H=22 |
* | The accuracy of the meter only. |
** | The accuracy of the meter only. The total error includes the meters and probe's accuracy. In the range 0 ÷ 100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8 °C, with Pt-1000A resistor: ±0.35 °C. |