SENSORS FOR OXYGEN IN THE AIR MEASUREMENTS
COG-3At, COG-3Atz


COG-3At is used for continuous measurements of oxygen concentration in the air in % and temperature. It co-operates with specific meters designed for co-operation with this sensor and continuous work (CO-421A)
COG-3Atz is used for est measurements of oxygen concentration in the air in % and temperature, equipped with multi pin connector. It co-operates with specific portable meters designed for co-operation with this sensor (CO-411A)

Characteristic features:



TECHNICAL DATA
SENSOR’S PARAMETERS
Parameter Model LS-25LF
Measuring range 0 ÷ 30%
Accuracy after calibration ± 1% of full scale
Work conditions Atmospheric pressure 811 hPa ÷ 1216 hPa
Temperature 5 ÷ 40°C
Relative humidity 10 ÷ 80% RH
without condensation
Response time ( T90%) about 15 s
External size of the housing Φ = 35 mm, L = 75 mm
Storing conditions Temperature -20°C ÷ 60°C
Relative humidity 0 ÷ 100% RH
Pressure 507 ÷ 1520 hPa

In the environment temperature down to -10°C sensor stops working but is not freezing, below -20°C it will be damaged.
The sensor may be directly mounted to the meter’s housing, or hanged in the chosen by the user place and connected with the meter with cable of chosen length.

INFLUENCE OF DIFFERENT GASES ON THE SENSOR (ACCORDING TO MANUFACTURER)
Type of the gas Concentration Influence on O2 measurement
Carbon dioxide 100,00% No influence
Hydrogen 100 ppm No influence
Hydrogen sulfide 50,3 ppm No influence
Nitric oxide 467 ppm No influence
Nitrogen dioxide 517 ppm +0,1 O2
Sulphur dioxide 46,6 ppm No influence
Ammonia 1,00% No influence

Temperature sensor: Pt-1000B platinum resistor
Accuracy in the range 0 ÷ 40°C: ±0,5° C