WATERPROOF OXYGEN METER CO-411A


The most important features




Detailed description

Designed for measurement of oxygen in the air in % and temperature.

Characteristic features:




Sensor characteristics:


  • The sensor is maintenance free, doesn’t require membrane nor electrolyte replacing.
  • Lifetime expectancy of the sensor – 5 years in standard conditions.
  • Sensor, after damage or use up, may be replaced by the user.
  • May be used in atmosphere with presence of other gases, details given in the table.
  • Standard cable length 1m, on order it may be changed.


  • TECHNICAL DATA
    SENSOR’S PARAMETERS
    Parameter Model LS-25LF
    Measuring range 0 ÷ 30% O2
    Accuracy after calibration ± 1% of full scale
    Work conditions Atmospheric pressure 811 hPa ÷ 1216 hPa
    Temperature 5 ÷ 40°C
    Relative humidity 10 ÷ 80% RH
    without condensation
    Response time ( T90%) about 15 s
    External size of the housing Φ = 35 mm, L = 75 mm
    Storing conditions Temperature -20°C ÷ 60°C
    Relative humidity 0 ÷ 100% RH
    Pressure 507 ÷ 1520 hPa

    In the environment temperature down to -10°C sensor stops working but is not freezing, below -20°C it will be damaged.

    INFLUENCE OF DIFFERENT GASES ON THE SENSOR (ACCORDING TO MANUFACTURER)
    Type of the gas Concentration Influence on O2 measurement
    Carbon dioxide 100,00% No influence
    Hydrogen 100 ppm No influence
    Hydrogen sulfide 50,3 ppm No influence
    Nitric oxide 467 ppm No influence
    Nitrogen dioxide 517 ppm +0,1 O2
    Sulphur dioxide 46,6 ppm No influence
    Ammonia 1,00% No influence

    Temperature sensor: Pt-1000B platinum resistor
    Accuracy in the range 0 ÷ 40°C: ±0,5° C

    TECHNICAL DATA OF THE METER
    Sensor calibration 1 point in the airr
    Accuracy (±1 digit) 0,1%*
    Memory 50 results
    Dimensions (mm) L=149 W=82 H=22
    Weight 220 g
    Power 9V battery
    optionally 12V / 100 mA power adapter
    Connector multi-pin
    * Final accuracy is a sum of meter and sensor errors.