WATERPROOF CONDUCTIVITY / SALINITY METER CC-411
The most important features
- Simple portable meter.
- Measures conductivity, salinity, TDS and temperature.
- The display enables simultaneous readout of the measured function and temperature.
- Conductivity calibration in 1 point by entering the constant K or in standard solution.
- Constant α coefficient (2%/°C).
- Memory of 50 results
- No PC connection.
Detailed description
CC-411 measures conductivity, salinity, TDS and temperature. May be used in the field and in the laboratory.Characteristic features:
- Measures conductivity in liquids with conductivity up to 200 mS/cm.
- The display enables simultaneous readout of the measured function and temperature.
- Low weight and small dimensions facilitate working in the field.
- Waterproof housing (IP-66) facilitates working in difficult conditions.
- 4 sub ranges switched automatically (autorange).
- Measures salinity with conversion to NaCl up to 170 g/l and KCl to 140 g/l.
- Converts conductivity into salinity according to real characteristic and not a constant coefficient.
- Determines estimated value of the total dissolved solids (TDS), variable coefficient introduced by user.
- Calibration by entering the constant K of the cell or by determining it with use of a standard solution.
- Automatic temperature compensation.
- Constant α coefficient (temperature coefficient) – 2%/°C. For precise measurements with temperature compensation it is recommended to use meters of 401 or 461 series.
- The reference temperature is 25 °C or 20 °C (e.g. for measurements in honey).
- Memory of 50 results
- Possibility to make measurements of electric admittance of tree seedlings (determining the vitality of seedlings with a special sensor).
- Automatic switch-off function saves the battery and protects it against discharging.
- Warranty for the meter: 24 months.
- The user's manual in English on a pen drive added to the set.
TECHNICAL DATA
Function | Conductivity | Salinity | Temperature |
---|---|---|---|
Range |
0 ÷ 199.9 mS/cm, autorange - 4 subranges |
NaCl 0 ÷ 170 g/l KCl 0 ÷ 140 g/l |
-50.0 ÷ 199.9 °C |
Accuracy (± 1 digit) | ±0.25%* | ±2%* | ±0.1°C** |
Temperature compensation | -5 ÷ 70 °C | -5 ÷ 70 °C | - |
α coefficient | constant 2%/°C | constant 2%/°C | - |
K constant | 0.010 ÷ 1,999 cm-1 | - | - |
Temperature sensor | Pt-1000B | ||
Power supply | 9 V battery, 12 V power adapter as option | ||
Dimensions (mm) | L = 149; W = 82; H = 22 | ||
Weight | 220 g |
* | The accuracy of the meter only. |
** | The accuracy of the meter only. The total error includes the meters and probe's accuracy. In the range 0 ÷100 °C the acceptable error of the probe with Pt-1000B resistor: ±0.8°C , with Pt-1000A resistor: ±0.35°C. |